TIS 166-2549 Figure A.2 Gauge for testing the contact of the live contact tube

TIS 166-2549 Figure A.2 gauge A/B/C for testing the contact of the live contact tube

TIS 166-2549 Figure A.2 Gauge for testing the contact of the live contact tube

Product No: SC-TIS166-2549-FA.2-A · FA.2-B · FA.2-C (gauges A, B and C)

Standard: TIS 166-2549 – Thai Industrial Standard for Plugs and Socket-Outlets for Household and Similar Purposes: Plugs and Socket-Outlets with Rated Voltage not Exceeding 250 V. Appendix A of the standard defines the gauges: Figures A.1 to A.5 check the dimensions of the socket-outlet, Figures A.6 and A.7 check the withdrawal force, and Figures A.8 to A.10 are the additional gauges for socket-outlets.

Application

Three gauges A, B and C check whether the live contact tube of a socket-outlet is reached at the correct depth. The shutters are kept open during the whole test and the contact condition is shown by an electronic indicator supplied at 40–50 V.

Gauge A fully inserted in any possible position of the socket opening, until plane X-X touches the engagement face of the socket-outlet: the indicator lamp shall not light.

Gauge B inserted in the socket-outlet: the indicator lamp shall light (the contact tube is reached).

Gauge C inserted in the socket-outlet: the indicator lamp shall not light (the contact tube is not reached too deep).

Construction

  • Three gauges A, B and C in corrosion-resistant stainless steel
  • Insulated sections on the body so that only the pin tip can signal a contact
  • Metal parts made of corrosion-resistant metal (stainless steel), as required by Figure A.2
  • Test leads for a 40–50 V electronic indicator
  • Engraved with the figure number and the gauge letter (A / B / C)

Key specification

Product No SC-TIS166-2549-FA.2-A / FA.2-B / FA.2-C
Reference TIS 166-2549 Figure A.2, clause 9.101
Gauge A Indicator lamp shall not light
Gauge B Indicator lamp shall light
Gauge C Indicator lamp shall not light
Test indicator 40–50 V electronic indicator

Product pictures

TIS 166-2549 Figure A.2 Gauge for testing the contact of the live contact tube - Gauges A, B and C - set view
Gauges A, B and C – set view
TIS 166-2549 Figure A.2 Gauge for testing the contact of the live contact tube - Gauges A, B and C - separated
Gauges A, B and C – separated
TIS 166-2549 Figure A.2 Gauge for testing the contact of the live contact tube - Gauges A, B and C - row view
Gauges A, B and C – row view
TIS 166-2549 Figure A.2 Gauge for testing the contact of the live contact tube - Gauges A, B and C - upright
Gauges A, B and C – upright
TIS 166-2549 Figure A.2 Gauge for testing the contact of the live contact tube - Gauges A and C - engraved bodies
Gauges A and C – engraved bodies

Standard drawing

TIS 166-2549 Figure A.2 standard drawing
Figure A.2 standard drawing – gauges A, B and C for the contact of the live contact tube (dimensions in mm)